Open original FDA record ↗Recall record
- FDA identifier
- Z-1870-2021
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- Jun 16, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
Source facts
- Action
- Siemens issued Urgent Field Safety Notice on May 4, 2021 domestically and abroad. The Urgent Medical Device Correction (UMDC), ASI21-03.A.US, was mailed to all affected US customers via FedEx. The Urgent Field Safety Notice (UFSN), ASI21-03.A.OUS, was issued electronically to all OUS Siemens sites with affected customers for distribution and regional reporting to their local competent authorities. Letter states reason for correction, health risk and action to take: If the Atellica CH 930 Analyzer in your laboratory posts event code 04 474 04 58, in the Operator Event Log, with the message, Photometer lamp intensity is low on at least one wavelength. Perform the Lamp Replacement routine in Operator diagnostics, stop the analyzer and replace the source lamp utilizing the Lamp Replacement routine, as prompted by the event message. Instructions for the Lamp Replacement routine are available by following the link in the event code message or by accessing the online help procedure, Replacing the Source Lamp. After the lamp has been replaced, any tests completed after the error message initially posted must be repeated. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Additional Information The Atellica Solution software will be updated to flag results for this unexpected lamp failure mode to address cases where a user may delay in responding to the lamp intensity event codes and alerts. Version 1.25.0, which is under development, will be available for installation on your analyzer soon. Please retain this letter with your laboratory records and forward this letter to those who may have received t
- Initiated
- 2021-05-04
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Croatia Cura¿ao, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, P.R. China Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Vatikancity, Vietnam.
- Recall number
- Z-1870-2021
- Recall status
- Open, Classified
- CDRH record ID
- 187504
- FDA record created
- 2021-06-16
- Recall event number
- 87953