Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.
Source facts
Action
Spine Frontier sent an Advisory Notice dated November 9, 2012, via email to all affected customers. Customers were advised that their FacetFuse Drill Guides need to be replaced.
Customers were instructed to write their number ( included in the email) on the outside of the package and provide us with the tracking information. A replacement part has been sent out to you.
If you have any further questions please call (978) 232-3990.
Initiated
2012-11-09
Root cause
Device Design
Terminated
2015-06-05
Distribution
US Distribution including Puerto Rico and the states of CO, KS, MA and TX.
Recall number
Z-1871-2013
Recall status
Terminated
CDRH record ID
117583
FDA classified
2013-08-02
Recall event number
64985