Open original FDA record ↗Recall record
- FDA identifier
- Z-1871-2016
- Company in FDA record
- Skytron, Div. The KMW Group, Inc
- FDA classification date
- May 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.
Source facts
- Action
- Skytron LLC initiated a voluntary recall of twenty-four monitor brackets used as a mounting point for display interfaces utilized in clinical applications for viewing purposes and to provide an articulating point of movement on the ceiling mounted radial arm suspension system of the Ergon 3 equipment pendant /booms via telephone calls to customers, and written notices sent via certified USPS mail on 04/01/2016. All potentially affected users will receive a field correction notification directly from Skytron, LLC along with the contact information of their local Skytron distributor. Skytron distributors must contact identified users within the universe of affected products to schedule a time replace the affected monitor bracket(s). Skytron will provide a replacement of the monitor bracket. The replacement monitor bracket(s) and installation must be conducted by you, the Skytron distributor. The affected monitor bracket(s) and Service Report indicating that the monitor bracket(s) was replaced must be returned to Skytron. Please contact Larry Perez, Vice President of Technical Services, at 616-656-1187 or lperez@skytron.us, or Thao Selleck, Complaint Resolution Analyst, at 616-656-6403 or tselleck@skytron.us with any questions. For after-hours support, please call 1-800-SKYTRON.
- Initiated
- 2016-03-28
- Root cause
- Process design
- Terminated
- 2017-07-25
- Distribution
- Worldwide distribution. US states: VA, VT, UT, GA, NJ, and IL; Japan.
- Recall number
- Z-1871-2016
- Recall status
- Terminated
- CDRH record ID
- 144784
- FDA classified
- 2016-05-31
- Recall event number
- 73704