Open original FDA record ↗Recall record
- FDA identifier
- Z-1872-2008
- Company in FDA record
- Biomet Microfixation, Inc.
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The outer foil pouch on the Lactosorb Plate system was not properly sealed. The product is labeled as sterile, but the lot was not sterilized. These devices are intended to be implanted into bone and are required to be sterile. A heavy bioburden would increase the potential for bone infection (osteomyelitis).
Source facts
- Action
- On 3/19/08 the firm issued an Urgent Medical Device Recall Notice to its consignees by certified letters with return receipt. All returned products will be held in quarantine until final disposition is determined. If you have any questions contact Rachel Osbeck at 1-800-874-7711 or (904) 741-4400 ext 448.
- Initiated
- 2008-03-19
- Root cause
- Packaging process control
- Terminated
- 2009-09-11
- Distribution
- Class II Recall - Worldwide Distribution --- USA including states of NC, MI, TX, GA, HI, OK, IN and CA, and countries of Denmark, Germany, Mexico, Italy, Brazil, Japan, Canada, South Africa, China and the Philippines.
- Recall number
- Z-1872-2008
- Recall status
- Terminated
- CDRH record ID
- 71474
- FDA classified
- 2008-09-16
- Recall event number
- 48461