Z-1873-2017: Laerdal Compact Suction Unit 4 under the following labels: 1) 300 ml (3.3 lbs) canister version - CATALO

Recall record Z-1873-2017 for Laerdal Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1873-2017
Company in FDA record
Laerdal Medical Corporation
FDA record creation date
Apr 20, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The On/Off functionality in the LCSU 4 is controlled by electronic components, one of which might be damaged.

Source facts

Action
On January 8, 2017 Laerdal Medical distributed Urgent Device Recall Notification Letters to their customers thru Stericycle. The documents were dispatched via standard UPS ground delivery. Laerdal Medical has discovered that the LCSU4 may turn on by itself when connected to a power source and/or may not be able to be turned off with the on/off button without removing the battery or external power source. You can continue to use your suction units until your replacement arrives by following the steps: Exception: Some LCSU 4 units have a RW label on the bottom. These units have been reworked by Laerdal Medical and are not affected by this Field Safety Notice, even if they have manufacturing dates within the 20150501 to 20161028 range. **Check On/Off functionality** You can run a quick test to identify if the unit has failing On/Off functionality: Turn unit on; Turn unit off again. If the unit cannot be turned off using the On/Off button, the unit contains a damaged component. The device should be taken out of service. The safety and effectiveness of the device is not affected by the failure of the On/Off button as long as it has power. Note that even if the unit passes this test, the problem may occur later. The actions below must be taken to ensure continued safe use of the LCSU 4 until you receive a replacement unit. **Actions to be taken to ensure continued safe use of the LCSU 4 until you receive a replacement unit** You can continue to use your LCSU 4 unit until you receive a replacement unit, but it is recommended to implement the following extra safety measures: Store the LCSU 4 unit without the battery connected so that it cannot turn on by itself. Perform routine checks of the On/Off functionality by turning the unit on and off using the On/Off button. All units with failing On/Off functionality should be removed from service. ** Actions to be taken if the On/Off functionality fails during clinical use** If the On/Off functionality fai
Initiated
2017-01-08
Root cause
Nonconforming Material/Component
Terminated
2018-05-25
Distribution
Nationwide
Recall number
Z-1873-2017
Recall status
Terminated
CDRH record ID
154075
FDA record created
2017-04-20
Recall event number
76744

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗