Reason stated in the source

Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended

Source facts

Action
On June 10, 2017, the recalling firm notified its Sales Representatives that effective June 10, 2017, the firm would no longer offer the temporal suture system (TSS). The firm requested that sales reps dispose of all marketing materials which contain details regarding the TSS. Sales reps who wanted to keep existing booklets in the field were asked to contact the firm to obtain adhesive labels to place over the discontinued product. On November 13, 2017, the recalling firm notified its Sales Representatives that the firm was temporarily suspending shipment of the Integrated Fixation System (IFS). The firm stated that all paper brochures were to be discarded, and asked sales reps to inform all those who received "Customize Skull and Craniofacial Implants" brochure and IFU. Revised marketing materials are available on the firm's website, www.kelyniam.com. If you have any questions, you may contact Dr. Mark V. Smith, Vice President of Kelyniam Global, Inc., at msmith@kelyniam.com, or by phone at (800) 280-8192 x702.
Initiated
2017-06-10
Root cause
Device Design
Terminated
2020-09-14
Distribution
US nationwide distribution.
Recall number
Z-1873-2018
Recall status
Terminated
CDRH record ID
163450
FDA record created
2018-05-14
Recall event number
79791

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗