Open original FDA record ↗Recall record
- FDA identifier
- Z-1874-2011
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Apr 15, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under some conditions during high doses procedures the cumulative air kerma display may reach its maximum capability and fail to display the accrued value.
Source facts
- Action
- An Important: Electronic Product Radiation Warning notification was sent to all consignees. The letters described the reason for the recall and identified the affected product. The system software will be modified to allow the shift of the decimal to one position to the right allowing for a display of 99999.9 mGy. This will allow for the retention of the resolution until such time as the system exceeds the 9999.99 mGy range. A Toshiba Service person will contact customers when the software is available. Installed systems with the current version of software ( version 4.23) will begin to receive the corrected software April 2011. Previous versions of the software will begin to receive corrected software June 2011. The software will be provided free of charge. Questions related to the letter should be directed to the Regulatory Affairs Department at 800-421-1968.
- Initiated
- 2010-12-21
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2013-03-18
- Distribution
- Nationwide Distribution
- Recall number
- Z-1874-2011
- Recall status
- Terminated
- CDRH record ID
- 98446
- FDA classified
- 2011-04-15
- Recall event number
- 58134