Z-1874-2017: Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of vent

Recall record Z-1874-2017 for Newport Medical Instruments Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1874-2017
Company in FDA record
Newport Medical Instruments Inc
FDA classification date
Apr 10, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Medtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.

Source facts

Action
Medtronic sent an Urgent Field Corrective Action Notice dated April 3, 2017, to all consignees via Federal Express or certified mail. The letter informed consignees of the remote potential for ventilator reset without accompanying alarm and the actions they should take. Customers were requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. Customers were also advised that a software service update will be available in May 2017 to address this issue. Medtronic issued a Press Release related to this voluntary field action on April 5, 2017. Consignees with questions were advised to contact the Technical Support Department immediately at 800-255-6774. For questions regarding this recall call 203-492-5415.
Initiated
2017-03-30
Root cause
Software in the Use Environment
Terminated
2020-07-18
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Canada, Australia, Brazil, Japan, Korea, Mexico, Bahrain, Egypt, Greece Hungary, Ireland, Israel, Jordan, Kazakhstan, Lithuania, Mauritius Namibia, Norway, Pakistan, Poland, Qatar, Romania, Russian Federation South Africa, Spain, Turkey, United Arab Emirates, United Kingdom Algeria, Armenia, Austria, Bahrain, Belarus, Belgium, Botswana Bulgaria, Cyprus, Czech Republic, Denmark, Egypt, Finland, France Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Kenya, Lebanon Lithuania, Macedonia, Malawi, Malta, Mauritius, Mozambique, Netherlands Norway, Oman, Poland, Portugal, Qatar, Republic of Moldova, Romania Saudi Arabia, Serbia, Slovakia, Sudan, Sweden, Switzerland, Tunisia Turkey, Uganda, Ukraine, United Kingdom, Yemen, and Zimbabwe
Recall number
Z-1874-2017
Recall status
Terminated
CDRH record ID
154521
FDA classified
2017-04-10
Recall event number
76891

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗