Skip to content

Public FDA record

Z-1874-2025: Catalog No. 8888160556; Umbilical Vessel Catheter 5 French dual lumen

Recall record Z-1874-2025 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1874-2025
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
May 29, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.

Source facts

Action
The firm started notifying customers on May 2, 2025 via Urgent Medical Device Product Recall letters. Customers were instructed to remove any devices currently in use. In addition, customers should review their inventory and segregate/quarantine all affected product. If product has been further distributed, the distributor should notify all downstream customers. Customers are asked to return an acknowledgement form whether they have affected product or not. Customers should contact the firm to arrange for return and credit or questions regarding suitable alternative products. For questions related to this recall, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Initiated
2025-05-02
Root cause
Package design/selection
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Canada, Chile, Colombia, Singapore, South Africa.
Recall number
Z-1874-2025
Recall status
Open, Classified
CDRH record ID
213653
FDA classified
2025-05-29
Recall event number
96783

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗