Open original FDA record ↗Recall record
- FDA identifier
- Z-1875-2009
- Company in FDA record
- Hill-Rom Manufacturing, Inc.
- FDA classification date
- Aug 3, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The devices may emit smoke from the blower box assembly.
Source facts
- Action
- Urgent Field Safety Notice letter dated May 11, 2009. Hill-Rom service centers were notified by letter on 05/13/2009. Action to be taken are as follows: Rental Customers; In the next 60 days, Hill Rom will be removing the Flexicair II and Flexicair MC3 low Airloss Therapy Units from service to correct the devices. During that time, Hill-Rom may provide an equivalent substitute therapy to accommodate the patients who continue to need a pressure reducing therapy device. If t his malfunction occurs before a substitution or a corrected device is provided, unplug the device, remove the patient from the surface and notify Hill-Rom immediately. Facility-Owned Customers: In the next 60 days, for all facility owned Flexicair II and Flexicair MC3 Low Airloss Therapy Units, a Hill-Rom service technician will either replace the affected components in the blower assembly or deliver an equivalent substitute unit from Hill-Rom's rental fleet. If this malfunction occurs before the device is corrected or a substitute provided, do as follows: unplug the device, remove the patient from the surface, and immediately notify Hill-Rom Technical Support using the attached listing their geographical region.
- Initiated
- 2009-05-12
- Root cause
- Equipment maintenance
- Terminated
- 2012-04-03
- Distribution
- Nationwide and Canada
- Recall number
- Z-1875-2009
- Recall status
- Terminated
- CDRH record ID
- 81939
- FDA classified
- 2009-08-03
- Recall event number
- 52032