Skip to content

Public FDA record

Z-1876-2024: Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm

Recall record Z-1876-2024 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1876-2024
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
May 28, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.

Source facts

Action
The recalling firm issued letters via email and first class mail on 2/26/2024. The recall letter informed the customer the recall was being conducted for specific items and lots of SubG Endotracheal Tube with Subglottic Suction due to receiving complaints that the inflation tube detached and/or tore from the main tube. There had also been reports indicating the suction pump is difficult to connect or detaches during use. The potential product impact was also provided. Affected product could be destroyed for credit and no product was to be returned. Required actions were provided: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. The link to the recall portal was provided. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form. The letter provided a link to their recall portal to access the response form for completion and view the specific products distributed to the respective customers. If the consignee is a distributor or has resold or transferred the product to another company or individual, they are to contact them and notify them of this recall communication and request they document and destroy any affected product and include the customer quantities in the distributor response form. The recalling firm issued a second recall letter dated 3/19/2024. The letter informed the consignee that the recalling firm was issuing a recall for specific items and lots of SubG Endotracheal Tube with Subglottic Suction with Medline Kits. If the consignee has the affected kits, they are to indicate the quantity
Initiated
2024-02-26
Root cause
Under Investigation by firm
Distribution
Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.
Recall number
Z-1876-2024
Recall status
Open, Classified
CDRH record ID
206447
FDA classified
2024-05-28
Recall event number
94147

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗