Skip to content

Public FDA record

Z-1877-2008: PANTA Nails, 10mm x 150 length, Newdeal, Lyon France; Integra, 311 Enterprise Drive, Plainsboro, NJ. The

Recall record Z-1877-2008 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1877-2008
Company in FDA record
Integra LifeSciences Corp.
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Certain PANTA Nails from one lot were incorrectly manufactured. The internal thread for certain PANTA Nails (Part Number 500050ND; Lot Number: E7XX) may present an insufficient depth and may not allow the engagement of the threaded part of the compression device. The risks presented include: an increase in surgery time, or the implantation of a nail less ideally suited for the patient's anatomy

Source facts

Action
On March 24, 2008, Recall Notification Letters and Recall Acknowledgement Forms were sent to all Integra Sales Specialists to confirm products currently in their inventory and to place inventory on hold with no further shipments to customers. If you have any questions contact Judy O'Grady at (609) 936-2311.
Initiated
2008-03-24
Root cause
Process control
Terminated
2008-09-16
Distribution
Class II Recall - Nationwide Distribution.
Recall number
Z-1877-2008
Recall status
Terminated
CDRH record ID
71736
FDA classified
2008-09-16
Recall event number
48621

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗