Skip to content

Public FDA record

Z-1877-2024: Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i

Recall record Z-1877-2024 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1877-2024
Company in FDA record
Getinge Usa Sales Inc
FDA classification date
May 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Source facts

Action
On May 1, 2024 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken by customer: Until further notice, do NOT use the Vaproizer with Sevoflurane provided by Poramal and Baxter Sevoflurane. Do NOT use the Vaporizer if it was ever used with Sevoflurane Piramal and/or Baxter Sevoflurane, even if it currently is used with AbbVie Ultane. Our records indicate that you have received the Vaporizer Sevoflurane, Maquet Filling having one or more of the serial numbers that are affected by this recall. 1. Please examine your inventory immediately to determine if you have any of the Vaporizer Sevoflurane, Maquet Filling with the product codes/serial numbers listed in this notice. The Product Code/REF Number and serial number (SN) are found on the label located underneath each vaporizer. Should you have any affected product, please remove from areas of use. 2. If you have ONLY used Vaporizer Product Code/REF Number 6682282, serial number 17003-23454 with AbbVie Ultane, you can continue to use the vaporizer, by certifying that you only use AbbVie Ultane on the Medical Device Recall Response Form. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 3. Persuant to the previous recall for the Vaporizer Product Code/REF Number 6886601, you can continue to use this vaporizer so long as it has ONLY been used with Abbvie Ultane. If you are uncertain whether you have previously used the vaporizer with Piramal Sevoflurane and/or Baxter Sevoflurane, discontinue use of the vaporizer. 4. If at any time, the agent in the vaporizer looks cloudy and yellowish, you should immediately stop use of the affected vaporizer. 5. If you have affected product you are entitled to a credit or replacement. Please contact Getinge Customer Service at 888-943-8872 (press option 2) to request a return authorization (RMA) and shipping instructions to return any affected product. Pack
Initiated
2024-05-01
Root cause
Under Investigation by firm
Distribution
Domestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.
Recall number
Z-1877-2024
Recall status
Open, Classified
CDRH record ID
207330
FDA classified
2024-05-31
Recall event number
94502

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.