Open original FDA record ↗Recall record
- FDA identifier
- Z-1880-2016
- Company in FDA record
- Ventlab LLC
- FDA classification date
- Jun 1, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.
Source facts
- Action
- Ventlab sent an URGENT: MEDICAL DEVICE PRODUCT REMOVAL notification dated March 14, 2016, on March 15, 2016, electronically to all affected customers. Customers were instructed to return the affected product and notify affected consignees of the recall. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this action, please call Ventlab at 1-844-635-5326, or email at PFA@ventlab.com.
- Initiated
- 2016-03-15
- Root cause
- Process control
- Terminated
- 2016-11-01
- Distribution
- Nationwide distribution to AZ, MS, NY, MO, FL, WI, TX, FL, and CA.
- Recall number
- Z-1880-2016
- Recall status
- Terminated
- CDRH record ID
- 144926
- FDA classified
- 2016-06-01
- Recall event number
- 73766