Open original FDA record ↗Recall record
- FDA identifier
- Z-1881-2017
- Company in FDA record
- Orthosoft, Inc. dba Zimmer CAS
- FDA classification date
- Apr 26, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal software ORTHOsoft-UniTkr-2.3.2.6, due to a calibration sequence crash.
Source facts
- Action
- Zimmer sent a Medical Device Field Notification letter dated December 6, 2011, to all affected customers. This is a retrospective report of a correction taken on 6 December 2011. The software issues were fixed in the new software version 2.3.3.0. Recall notices were distributed on December 6, 2011. Telephone: 18663367846 (US & Canada) or 5143958883 (International), Email: cassupport@zimmercas.com Customers with questions should call 1-866-336-7846 (US & Canada) 514-395-8883 (International). For questions regarding this recall call 574-372-4487.
- Initiated
- 2011-10-06
- Root cause
- Software design
- Terminated
- 2017-05-08
- Distribution
- Worldwide Distribution - US including AZ CA CO FL GA IL IN KS MD MI MN MS OH OR PA SC SD TX VA WA and WI Internationally to Canada, Australia Austria Colombia Denmark France Germany Korea New Zealand Russia South Africa and Thailand
- Recall number
- Z-1881-2017
- Recall status
- Terminated
- CDRH record ID
- 154874
- FDA classified
- 2017-04-26
- Recall event number
- 75614