- Recall classification
- Class II
Reason stated in the source
There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
Source facts
- Action
- On 4/22/24, Cadwell Industries, Inc. issued a "Urgent: Medical Device Recall" notification to affected consignees via Email: 1. Cease use of the device immediately and return all identified devices to Cadwell Industries, Inc. 2. Please complete the attached response form and return via email to quality@cadwell.com. 3. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred the IOMAX Cortical Module to. 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
- Initiated
- 2024-04-23
- Root cause
- Process design
- Distribution
- Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.
- Recall number
- Z-1882-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207342
- FDA classified
- 2024-05-22
- Recall event number
- 94504