Open original FDA record ↗Recall record
- FDA identifier
- Z-1883-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 25, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Not properly labeling the manual switch for disabling positive beam limitation (PBL) located on the overhead tube support user interface as required
Source facts
- Action
- GE Field Engineer will visit each site to implement the correct label. Contact GE Healthcare at 1-262-544-3894 for assistance.
- Initiated
- 2008-01-02
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2011-12-10
- Distribution
- AL, CA, CO, CT, DE, FL, GA, IA,, IL,, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI,WV and Washington D.C. OUS to include: Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Malta, Martinique (French), Netherlands, Norway, Palestine, Portugal, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, and Yugoslavia,
- Recall number
- Z-1883-2008
- Recall status
- Terminated
- CDRH record ID
- 73598
- FDA classified
- 2008-09-25
- Recall event number
- 49487