Skip to content

Public FDA record

Z-1883-2016: Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIM

Recall record Z-1883-2016 for Greatbatch Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1883-2016
Company in FDA record
Greatbatch Medical
FDA record creation date
Jun 1, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Source facts

Action
Consignees were sent on 4/26/2016 a Greatbatch "Urgent Medical Device Recall" letter dated April 26, 2016. The letter described the product involved in the recall, the reason, and the risk to health. The letter advised consignees to identity and quarantine the product and to complete and return the Filed Recall Response Form. For questions contact your local Greatbatch Medical representative, at 1-763-951-8235, or e-mail FieldActionCenter@Greatbatch.com.
Initiated
2016-04-26
Root cause
Device Design
Terminated
2018-06-07
Distribution
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Recall number
Z-1883-2016
Recall status
Terminated
CDRH record ID
145758
FDA record created
2016-06-01
Recall event number
73948

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗