Open original FDA record ↗Recall record
- FDA identifier
- Z-1883-2016
- Company in FDA record
- Greatbatch Medical
- FDA record creation date
- Jun 1, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.
Source facts
- Action
- Consignees were sent on 4/26/2016 a Greatbatch "Urgent Medical Device Recall" letter dated April 26, 2016. The letter described the product involved in the recall, the reason, and the risk to health. The letter advised consignees to identity and quarantine the product and to complete and return the Filed Recall Response Form. For questions contact your local Greatbatch Medical representative, at 1-763-951-8235, or e-mail FieldActionCenter@Greatbatch.com.
- Initiated
- 2016-04-26
- Root cause
- Device Design
- Terminated
- 2018-06-07
- Distribution
- US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
- Recall number
- Z-1883-2016
- Recall status
- Terminated
- CDRH record ID
- 145758
- FDA record created
- 2016-06-01
- Recall event number
- 73948