Open original FDA record ↗Recall record
- FDA identifier
- Z-1883-2017
- Company in FDA record
- CareFusion 303, Inc.
- FDA record creation date
- Apr 27, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
CareFusion is recalling the SmartSite Add-On Bag Device due to reports of separation and/or leakages between the spike port and the drip chamber spike. A separation or leak can cause a delay of infusion, an interruption of infusion, exposure to medication or hazardous infusates, or underinfusion.
Source facts
- Action
- CareFusion sent an Urgent Medical Device Recall Notification letter dated November 2016, to inform their customers that CareFusion is recalling the SmartSite Add-On Bag Access Device Model 10013365 because of reports of leakages between the spike port and the drip chamber spike. The letter informs the customers of the lot numbers, potential risk and immediate actions to be taken. Customers with questions are instructed to contact CareFusion Support Center at (888) 562-6018 for recall related questions. Customers with adverse event report questions are instructed to contact Customer Advocacy at (888) 812-3266 or by email at customerfeedback@carefusion.com. Customers are instructed to complete and return the enclosed mandatory customer response card.
- Initiated
- 2016-11-17
- Root cause
- Other
- Terminated
- 2018-10-29
- Distribution
- US Distribution
- Recall number
- Z-1883-2017
- Recall status
- Terminated
- CDRH record ID
- 149125
- FDA record created
- 2017-04-27
- Recall event number
- 75013