Open original FDA record ↗Recall record
- FDA identifier
- Z-1884-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 4, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Device function failure--When performing a TEE exam with the transducer, CW (Continuous Wave) mode is not functional. Due to work flow, this may not be apparent until the transducer is inserted into the patient's esophagus.
Source facts
- Action
- Siemens sent a Customer Safety Advisory Notification letter on April 7, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the notice on to all those within their organization who need to be aware of the issue until the corrective action is completed. For questions regarding this recall call 650-969-9112.
- Initiated
- 2009-04-07
- Root cause
- Software design
- Terminated
- 2011-04-05
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Canada, Switzerland, China, Czech Republic, Germany, Denmark, Great Britain, Hong Kong, Indonesia, India, Italy, South Korea, Norway, Turkmenistan, and Taiwan
- Recall number
- Z-1884-2011
- Recall status
- Terminated
- CDRH record ID
- 84089
- FDA classified
- 2011-04-04
- Recall event number
- 52820