AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Source facts
Action
FujiFilm issued Urgent Medical Device Recall Letter via email on 4/11/24. Letter states reason for recall, health risk and action to take:
Please complete and return the Field Action Verification Form attached on page 3 of this communication.
Responding with the information requested is essential for ensuring appropriate action is taken.
Please accept the replacement of AC adapter that is implemented the countermeasure.
Please contact our field action coordinator, Kotei Aoki, via email at
HCUSRegulatoryAffairs@fujifilm.com, if you have any further questions regarding this field action.
Initiated
2024-04-11
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-1884-2024
Recall status
Open, Classified
CDRH record ID
207222
FDA classified
2024-05-22
Recall event number
94441