Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The collimator could fall from the x-ray stand, due to loosening screws, and strike a patient or operator underneath.
Source facts
Action
On 2/7/08, the firm issued an URGENT-DEVICE CORRECTION letter via certified mail to their customers. The letter explains the issue and states that the firm will upgrade the collimator by installing a safety plate to prevent the collimator form dislodging from the stand should the screws become lose. The letter also states, "To avoid the potential risk of the collimator falling from the stand, please test the collimator for being loose. Rock the collimator forwards and backwards by grabbing the sides and gently shaking the unit. Please contact Philips immediately if the collimator is loose and do not use the system until Philips can service the unit." The service representatives will visit the sites and install a safety plate to prevent the collimator from falling should the securing screws become loose.
Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
Initiated
2008-02-07
Root cause
Device Design
Terminated
2011-03-08
Distribution
Units were distributed nationwide to medical centers and hospitals, including military and VA hospitals.
Recall number
Z-1886-2008
Recall status
Terminated
CDRH record ID
67007
FDA classified
2008-09-16
Recall event number
46216
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.