Z-1886-2020: BodyGuard Microset For Use with the BodyGuard Infusion Pump *** Tubing: 110 in. (280 cm) - Blue Stripe -

Recall record Z-1886-2020 for CME America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1886-2020
Company in FDA record
CME America, LLC
FDA record creation date
May 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The filter installed onto the tube set may have the incorrect pore size.

Source facts

Action
On 12/20/2019, the firm contacted its distributors via phone call to request that the affected lots not be further distributed pending the Recalling Firm's investigation and assessment for recall. On 01/15/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification to customers via email and FedEx informing them that the filter installed onto the tube set may have the incorrect pore size: -For the lot 90829, the specified size on the label is 0.2 microns. A portion of the tube sets were incorrectly packaged and have a pore size of 1.2 microns rather than 0.2 microns. -For the lot 16201, the specified size on the label is 1.2 microns. A portion of the tube sets were incorrectly packaged and have a pore size of 0.2 microns rather than 1.2 microns. Customer are instructed to take the following actions: 1. Immediately review their inventory for Catalog #A120-160XSFK/lot number 90829 and Catalog # A120-112XSFK/lot number 16201. Discard all remaining inventory of the tube sets per their facilitys guidelines. 2. Share this Urgent Medical Device Recall notice with all users of the product within their facility to ensure awareness of this action. 3. Complete the attached Customer Response Form and return it to the contact noted on the form regardless of whether you have any affected material or not so that we may acknowledge your receipt of this notification. Indicate on the response form the quantities of affected product remaining in inventory and confirm that this inventory has been discarded. Credit will be issued to customers for all discarded product. 4. Report any adverse health consequences experienced with the use of this product to the Recalling Firm. 5. Contact the Recalling Firm at 877-263-0111 (toll free) or 303-939-4545, Monday thru Friday from 9:00am to 5:00pm U.S. Mountain Time for a list of alternate sets that are available.
Initiated
2020-01-14
Root cause
Process control
Terminated
2021-05-27
Distribution
US Nationwide distribution including in the states of AK, CA, IL, KS, MA, MO, and SD. No O.U.S.
Recall number
Z-1886-2020
Recall status
Terminated
CDRH record ID
180276
FDA record created
2020-05-04
Recall event number
85206

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗