Z-1887-2024: IMMULITE 2000 Anti-TG Ab Catalog # L2KTG6 (600 Tests), SMN 10381655, System, Test, Thyroid Autoantibody

Recall record Z-1887-2024 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1887-2024
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
May 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential for falsely elevated patient sample results when using impacted lots (785, 790, 791, and 793) of IMMULITE 2000 Anti-TG Ab.

Source facts

Action
On April23, 2024, Urgent Medical Device Correction letters were sent to customers. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Perform the following: Discontinue use of and discard the kit lots listed in the Products Table. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product.
Initiated
2024-04-23
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution including in the states of AL, CA, DC, FL, HI, ID, IN, KS, KY MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, PR, TX, WA WI and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Finland, France, Germany, Greece Guatemala, Honduras, Hong Kong, Hungary, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Norway, P.R. China, Pakistan, Paraguay, Peru, Poland, Portugal, Republic South Korea, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, Ukraine, United Kingdom, Uruguay, Vietnam.
Recall number
Z-1887-2024
Recall status
Open, Classified
CDRH record ID
207317
FDA classified
2024-05-23
Recall event number
94495

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗