Open original FDA record ↗Recall record
- FDA identifier
- Z-1888-2020
- Company in FDA record
- Inpeco S.A.
- FDA record creation date
- May 5, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Source facts
- Action
- The firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.
- Initiated
- 2020-04-07
- Root cause
- Software design (manufacturing process)
- Distribution
- Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
- Recall number
- Z-1888-2020
- Recall status
- Open, Classified
- CDRH record ID
- 180858
- FDA record created
- 2020-05-05
- Recall event number
- 85429