Skip to content

Public FDA record

Z-1889-2020: Siemens Aptio Automation Centrifuge Module, tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware

Recall record Z-1889-2020 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1889-2020
Company in FDA record
Inpeco S.A.
FDA record creation date
May 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

Source facts

Action
The firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.
Initiated
2020-04-07
Root cause
Software design (manufacturing process)
Distribution
Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
Recall number
Z-1889-2020
Recall status
Open, Classified
CDRH record ID
180873
FDA record created
2020-05-05
Recall event number
85429

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗