Z-1889-2024: Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog N

Recall record Z-1889-2024 for Channel Medsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1889-2024
Company in FDA record
Channel Medsystems, Inc.
FDA classification date
May 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Error code 003 was not listed in certain rows of the Device Instructions for Use Troubleshooting section (i.e., Table 22. Summary of LCD Messages and Error Codes).

Source facts

Action
On 05/03/2024, the firm sent an "URGENT FIELD SAFETY NOTICE" via email to customers informing them that Channel Medsystems is updating its troubleshooting instructions to further inform healthcare professionals on the safe use and troubleshooting of the Cerene Device. the IFU has been updated to clarify that error code 003 can occur at any time (not only before, but also during, or after cryoablation treatment) and to instruct the user to continue to follow the correctly displayed Cerene Device LCD message and corresponding IFU troubleshooting instructions specifically, Uterus Partially Treated, End procedure, Do NOT re-treat or Uterus Treated End procedure Do NOT re-treat as the Treatment Status & Next Steps matching the corresponding displayed Cerene Device LCD screen message/error code 003. Customer are instructed to: 1. Follow your facility s established procedures for handling updated labeling/documents, such as immediately providing a copy of this notice to relevant healthcare professionals and notifying them of the updated IFU (Troubleshooting) and its online availability. 2. No product is being recalled and you are not required to return devices or IFUs to Channel Medsystems. 3. Complete and return the acknowledgment form at the end of this urgent field safety notice within one month of receipt. Questions or assistance - contact the firm at 510-338-9301 or email safetynotice@cerene.com during Office Hours: Monday through Friday. 8:00 AM to 5:00 PM, Pacific Time.
Initiated
2024-05-03
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin and the country of Netherlands.
Recall number
Z-1889-2024
Recall status
Open, Classified
CDRH record ID
207337
FDA classified
2024-05-23
Recall event number
94506

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗