Open original FDA record ↗Recall record
- FDA identifier
- Z-1890-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 17, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner.
Source facts
- Action
- Consignees were sent a "GE Healthcare Product Safety Notification" letter dated 9/14/07. The letter was addressed to Hospital Administrators, Managers of Radiology/Cardiology and Radiologists/Cardiologists. The letter described the safety issue, affected product, safety instructions, product correction and contact information. FMI 60745 and 60746 was communicated directly to the GE Healthcare service engineers who provide the modification at the customer site. For assistance, contact GE Healthcare at 1-262-544-3894.
- Initiated
- 2007-12-14
- Root cause
- Software design
- Terminated
- 2012-04-12
- Distribution
- AL, CA, CO,CT, AZ, FL, GA, LA, MS, MT, IA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NY, NV, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PUERTO RICO. AUSTRALIA, CHINA, HONG KONG, INDIA, JAPAN, SOUTH KOREA, NEW ZEALAND, TAIWAN, SWITZERLAND, GERMANY, DENMARK, SPAIN, FRANCE, GREAT BRITAIN, IRELAND, IRAN, ITALY, KUWAIT, LEBANON, NETHERLANDS, NORWAY, RUSSIA, SAUDI ARABIA, TURKEY, SOUTH AFRICA, BRAZIL, CANADA, and MEXICO.
- Recall number
- Z-1890-2008
- Recall status
- Terminated
- CDRH record ID
- 70798
- FDA classified
- 2008-09-17
- Recall event number
- 48073