Z-1891-2010: IMMULITE 2000/2500 Chemiluminescent Substrate Module For the use in the routine operation of the IMMULIT

Recall record Z-1891-2010 for Siemens Healthcare Diagnostics Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1891-2010
Company in FDA record
Siemens Healthcare Diagnostics Inc.
FDA classification date
Jun 24, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

These substrate lots have shown a tendency to introduce a shift in both control and patient results with multiple methods.

Source facts

Action
An Urgent Field Safety Notice dated May 2009 was sent to consignees of record for the previously mentioned lots. The Notice advised the following: These substrate lots have shown a tendency to introduce a shift in both control and patient results with multiple methods. Actions If you are using one of these lots with your instrument, please take the following actions: 1. Upon receipt of a non-affected substrate lot, discontinue using IMMULITE 2000/2500 Chemiluminescent Substrate. Prior to receiving a non-affected substrate lot, you can continue to use the affected substrate if you follow the Interim Procedure. 2. Discard all bottles of the affected Substrate lots remaining in your inventory once non-affected Substrate lots are available. 3. The first time ONLY that you switch from one of the affected substrate lots in Table 1 to a non-affected lot, you should follow the procedure and steps below during instrument Startup: a. Empty the substrate reservoir completely before adding the new lot. b. Prime the instrument to cleanse the lines of the previous lot of substrate. c. Add the new lot of substrate to the reservoir and prime the instrument again.* d. Adjust all methods run on that instrument. *Note: Prime the substrate line as instructed on Page 83, Priming the Substrate Probe, of the IMMULITE 2000/2500 Operators Manual Rev. A. 4. Discuss the contents of this letter with your laboratory director. 5. Complete the Completion Notification Form that accompanies this notice and return it as directed on the form. Submitting this form indicates that you have received the information in the Urgent Field Safety Notice and have either discarded all lots of the affected IMMULITE 2000/2500 Substrate, or are following the Interim Procedure while awaiting the receipt of a non-affected Substrate lot and will discard the affected lots as soon as the non-affected lots are received in your laboratory. Interim Procedure Siemens recognizes that you
Initiated
2009-05-15
Root cause
Device Design
Terminated
2011-01-18
Distribution
Nationwide Distribution
Recall number
Z-1891-2010
Recall status
Terminated
CDRH record ID
82515
FDA classified
2010-06-24
Recall event number
52206

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗