Open original FDA record ↗Recall record
- FDA identifier
- Z-1891-2018
- Company in FDA record
- Volcano Corporation
- FDA classification date
- May 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.
Source facts
- Action
- Customers were notified on 03/21/2018. U.S. customers were notified via hand-delivered Urgent Medical Device Recall letters; foreign customer letter was mailed. The letter identified the affected device and the reason for the recall. Instructions included to stop usage of the device and to call remote support at 1-800-228-4728, to schedule a visit to replace the affected device. Customers were instructed to complete and return the customer response form or to provide the form to Philips representatives. Customers were also instructed to provide copies of the recall notification to all personnel within their organization who handled this device.
- Initiated
- 2018-03-22
- Root cause
- Other
- Terminated
- 2021-04-28
- Distribution
- Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.
- Recall number
- Z-1891-2018
- Recall status
- Terminated
- CDRH record ID
- 163160
- FDA classified
- 2018-05-17
- Recall event number
- 79715