Skip to content

Public FDA record

Z-1891-2020: daVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref: 48360T-08 , UDI: 00886874113851, Rx only,

Recall record Z-1891-2020 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1891-2020
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
May 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.

Source facts

Action
On July 1, 2019 , the firm sent an "Urgent Medical Device Recall" notification to affected customers concer the da Vinci Xi SureForm 60 Stapler Reload. In addition to informing customer's about the recalled product, the firm asked customers to take the following actions: 1. Intuitive Surgical requests customers locate the affected Stapler Reloads at each site and return them. 2. Inform necessary personnel when corrective action has been completed. 3. If customer has distributed any affected product to other sites, please forward this notice to all related parties. 4. If available, please log into the da Vinci Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 5. Complete the attached Acknowledgement Form and return it via fax to Intuitive Surgical as instructed on the form, or complete using the da Vinci online resource tool.
Initiated
2019-07-01
Root cause
Employee error
Distribution
Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.
Recall number
Z-1891-2020
Recall status
Open, Classified
CDRH record ID
180397
FDA record created
2020-05-05
Recall event number
85256

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗