Open original FDA record ↗Recall record
- FDA identifier
- Z-1892-2010
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jun 24, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Siemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study instance UIDs.
Source facts
- Action
- The recalling firm sent a Customer Safety Advisory Notice, dated 04-06-2010, to customers explaining when the malfunction occurs and the potential risks, steps to avoid potential risk of the issue by ensuring that the RIS works according to DICOM Standard and does not create double study instance UIDs. Siemens Service Reps will contact the consignees in a timely manner to check if there are symptoms for the reported issues. The issue will be resolved with a Software-Update for each affected version, except for version VB20. All customers on VB20 will be upgraded.
- Initiated
- 2010-04-06
- Root cause
- Device Design
- Terminated
- 2012-07-05
- Distribution
- Nationwide Distribution -- Including CT, NM, & WI.
- Recall number
- Z-1892-2010
- Recall status
- Terminated
- CDRH record ID
- 91460
- FDA classified
- 2010-06-24
- Recall event number
- 55651