Z-1892-2019: PROBEAT-V, software versions 4.8.1S, and 5.1M Radiation therapy system.

Recall record Z-1892-2019 for Hitachi America, Ltd., Power Systems Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1892-2019
Company in FDA record
Hitachi America, Ltd., Power Systems Division
FDA record creation date
Jun 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for a discrepant target position when using 3D3D matching mode in PIAS (Positioning Image Analysis System) software installed in the PROBEAT-V, proton therapy systems.

Source facts

Action
The firm informed affected user sites of the recall beginning on April 15, 2019, via an Engineering Document. The firm distributed updated Urgent Medical Device Recall letters in June 2019. The firm stated that the PIAS software would be replaced with the corrected version. Until the software was corrected, users were asked to take the following interim countermeasures: -Limit the number of treatment planning CT slices to odd numbers only. - Create treatment plan only in Head First condition for new patients. - Use the beam limiting device in full open in the lateral direction (same as before). Customers were asked to complete and return the customer response form to the firm. The firm stated that the software correction will be implemented from June 17 - 29, 2019.
Initiated
2019-04-15
Root cause
Software design
Terminated
2020-04-24
Distribution
Distributed to accounts located in AZ, MN, and TN. Foreign distribution to Japan.
Recall number
Z-1892-2019
Recall status
Terminated
CDRH record ID
173428
FDA record created
2019-06-28
Recall event number
83056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗