- FDA identifier
- Z-1892-2025
- Company in FDA record
- Civco Medical Instruments Co. Inc.
- FDA classification date
- May 30, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Source facts
- Action
- An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/1/25 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any BX2 Needle Guides and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any BX2 Needle Guides remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to distorder@civco.com. Type of Action by the Company: This notice is being communicated to all customers who have purchased a BX2 Needle Guide from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level. PLEASE SEND COMPLETED RESPONSE FORM to email to distorder@civco.com. If you have any questions, call 319-248-6502 or email: James.Leong@civco.com.
- Initiated
- 2025-05-01
- Root cause
- Process design
- Distribution
- OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark
- Recall number
- Z-1892-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213714
- FDA classified
- 2025-05-30
- Recall event number
- 96814