Z-1892-2025: BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with th

Recall record Z-1892-2025 for Civco Medical Instruments Co. Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1892-2025
Company in FDA record
Civco Medical Instruments Co. Inc.
FDA classification date
May 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/1/25 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any BX2 Needle Guides and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any BX2 Needle Guides remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to distorder@civco.com. Type of Action by the Company: This notice is being communicated to all customers who have purchased a BX2 Needle Guide from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level. PLEASE SEND COMPLETED RESPONSE FORM to email to distorder@civco.com. If you have any questions, call 319-248-6502 or email: James.Leong@civco.com.
Initiated
2025-05-01
Root cause
Process design
Distribution
OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark
Recall number
Z-1892-2025
Recall status
Open, Classified
CDRH record ID
213714
FDA classified
2025-05-30
Recall event number
96814

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗