A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.
Source facts
Action
The firm initiated the recall by letter on 12/04/2017. The letter identified the affected product, problem and the actions to be taken. The firm requested destruction of the recalled product.
Initiated
2017-12-04
Root cause
Mixed-up of materials/components
Terminated
2023-05-30
Distribution
Worldwide Distribution and US Nationwide Distribution
Recall number
Z-1893-2018
Recall status
Terminated
CDRH record ID
163764
FDA record created
2018-05-18
Recall event number
79857