Open original FDA record ↗Recall record
- FDA identifier
- Z-1894-2010
- Company in FDA record
- Hospira Inc
- FDA classification date
- Jul 2, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Hospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation.
Source facts
- Action
- Hospira issued a "Urgent: Medical Device Recall Notification" dated June 11, 2010 to update a previously issued Clinical Bulletin. Customers were informed of the issue and instructed on proper mitigation steps to further avoid the issue associated with the affected product. Consignees were asked to complete and return the Reply Form. For further information, contact Hospira Technical Support Operations at 1-800-332-4002, 6AM to 4PM PST, Monday through Friday.
- Initiated
- 2010-04-09
- Root cause
- Device Design
- Terminated
- 2013-12-03
- Distribution
- Worldwide Distribution -- United States (AL, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA and WA), and Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and United Arab Emirates.
- Recall number
- Z-1894-2010
- Recall status
- Terminated
- CDRH record ID
- 92057
- FDA classified
- 2010-07-02
- Recall event number
- 55854