Z-1894-2018: iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine

Recall record Z-1894-2018 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1894-2018
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Apr 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.

Source facts

Action
The recalling firm, Beckman Coulter/IRIS, sent an "URGENT MEDICAL DEVICE RECALL" letter dated dated 3/26/2018 via email on 3/29/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Follow the on-board strip stability claims per Table 1 to determine the frequency at which the strips need to be changed. -Whenever the Strip Provider Module is accessed (i.e., opened or when loading strips), remove and discard the remaining on-board strips and desiccant before replenishing. -It is suggested that for the tests performed since the installation of the iChemVELOCITY/SPM housing assembly, consult with your Laboratory Director to determine whether a retrospective review of results is clinically warranted. -Complete and return the CUSTOMER RESPONSE FORM within 10 days by email to: Regulatory.Notifications@beckman.com, Please include FA-33476 iChemVELOCITY Automated Urine Chemistry System in e-mail subject field or by mail to:Beckman Coulter, Inc., Attn: Regulatory Affairs Mail Code 31-B06 11800 S.W. 147th Avenue, P O BOX 169015, Miami, FL 33116-9015. -Beckman Coulter will contact you by phone to schedule a site visit for replacing the SPM on your instrument. -Beckman Coulter will be implementing process improvements to prevent this issue from recurring. If you have any questions regarding this notice, please contact: - From our website: http://www.beckmancoulter.com - By phone: call 800-854-3633 in the United States and Canada. -Outside the United States and Canada, contact your local Beckman Coulter representative.
Initiated
2018-03-29
Root cause
Process control
Terminated
2019-06-27
Distribution
US Distribution to states of: AL, IA, GA, and FL.
Recall number
Z-1894-2018
Recall status
Terminated
CDRH record ID
163822
FDA classified
2018-04-27
Recall event number
79895

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗