Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Bright artifacts may appear on a patient image from a previous exposure, which might lead to a misdiagnosis.
Source facts
Action
On 4/08/08, the firm notified customers by letter, sent by certified mail. The letter explained the recall and advised customers of the following: "If direct radiation reaches the detector, the collimation should be not too tight, such that the area of direct radiation is larger than 30 mm x 30 mm. (1.2 inch by 1.2 inch). The reason is that a minimum area is needed to detect and prevent memory artifact... All possibly affected products in the field will be upgraded with FCO 71200026 to resolve the issue."
Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if you have questions.
Initiated
2008-04-08
Root cause
Software design
Terminated
2011-06-07
Distribution
Nationwide Distribution -- including states of SC, MN, CA, GA, TX, NY, OR and UT.
Recall number
Z-1895-2008
Recall status
Terminated
CDRH record ID
69721
FDA classified
2008-09-16
Recall event number
47539