Open original FDA record ↗Recall record
- FDA identifier
- Z-1895-2017
- Company in FDA record
- Richard-Allan Scientific Company
- FDA classification date
- Apr 17, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Methylene Blue, Solution #3 in the Kwik Diff Kit was not Methylene Blue but Crystal Violet instead.
Source facts
- Action
- On 3/15/2017, URGENT MEDICAL DEVICE RECALL letters were sent to the affected consignees. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER Please contact Richard Allan Scientific for directions on safe disposal or return if customers have affected lots. All other lots of the products, or similar products manufactured by Richard Allan Scientific, are unaffected. Upon identification of the affected lot, please notify Richard Allan Scientific of the quantity and the preferred disposal/return method. If product is being disposed of, evidence of disposal is required to be sent to Richard Allan Scientific. If affected product has been used with no issues, please notify Richard Allan Scientific of the quantity that is unavailable for return or disposal. Please use the attached Recall Return Response Acknowledgement & Receipt Form.
- Initiated
- 2017-03-15
- Root cause
- Mixed-up of materials/components
- Terminated
- 2017-07-27
- Distribution
- Worldwide Distribution -- US: CA, CO, CT, DE, GA, HI, IA, IL, KS, MA, MS, NC, NH, NM, NY, OH, PA, PR, RI, UT, VA, VT, WA, and WV; and, the countries of AE, BH, CA, CZ, FR, GB, HK, SA, and TR.
- Recall number
- Z-1895-2017
- Recall status
- Terminated
- CDRH record ID
- 154823
- FDA classified
- 2017-04-17
- Recall event number
- 77011