Open original FDA record ↗Recall record
- FDA identifier
- Z-1896-2008
- Company in FDA record
- Curlin Medical Llc
- FDA classification date
- Sep 22, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Over-delivery/free-flow involving Curlin ambulatory peristaltic multi-therapy infusion pumps where the pumping chamber door (platen) had become deformed in a manner that permitted gravity flow.
Source facts
- Action
- On May 30, 2008, Curlin Medical began distribution of a "Voluntary Field Correction Advisory Notice" to all customers in the United States via USPS Registered Mail. The Notice informed customers of the problem and gave the customers (1) Guidelines on testing pump volumetric accuracy, (2) Updated user manuals to elevate the "caution" statement related to dropped pumps to a "warning" statement, and (3) A warning label to be located on the pump: "Warning: Impact may cause damage. If dropped, pump must be checked for accuracy prior to use". A toll free number (888-287-5999) was given for customers experiencing issues that require immediate assistance.
- Initiated
- 2008-05-30
- Root cause
- Other
- Terminated
- 2009-06-11
- Distribution
- Worldwide Distribution --- including USA and countries of Australia, Croatia, Germany, Italy, New Zealand, and South Africa.
- Recall number
- Z-1896-2008
- Recall status
- Terminated
- CDRH record ID
- 71773
- FDA classified
- 2008-09-22
- Recall event number
- 48639