Skip to content

Public FDA record

Z-1896-2021: OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical

Recall record Z-1896-2021 for Ossur Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1896-2021
Company in FDA record
Ossur Americas
FDA record creation date
Jun 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.

Source facts

Action
On 06/01/2021, the Recalling Firm emailed or called its customers and provided an "URGENT MEDICAL DEVICE RECALL" letter informing them that it had identified that cervical collars may deform under certain conditions, possibly leading to reduced immobilization of the cervical spine. The likelihood of deformation is higher when in use at higher height configurations. Further, the risk of deformation increases with the number of days the device is worn. Customers are instructed to: -Examine their inventory, quarantine products subject to the recall, and return the Customer Acknowledgement Form to the email address listed on the form. A return authorization will then be provided to return the product to the Recalling Firm. Additional actions to take: 1. Pass this notice to those who need to be aware within the customer's organization or to any organization where the potentially affected devices has been transferred. 2. Maintain awareness on this notice for an appropriate period. 3. If the affected products have been further distributed, identify the customers, and notify them at once of this recall. The Recalling Firm recommends including a copy of this recall letter. 4. The Recalling Firm recommends healthcare professionals perform a detailed inspection of the device to verify lot number, and to ensure proper fit and structural stability for any patients current wearing impacted product. 5. If any customers are currently wearing a product, the packaging label should be inspected for the lot number (Note: The number inside the collar is not the lot number; that is the device serial number). " If the lot number is outside of the affected range, no actions need to be taken. " If the lot number is inside the affected range, healthcare professionals are advised to assess whether a removal and replacement of impacted product currently in use is warranted on a case-by-case basis. " If the packaging label/lot number is not available, healthcare professionals a
Initiated
2021-06-01
Root cause
Device Design
Terminated
2023-06-22
Distribution
U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Not provided.
Recall number
Z-1896-2021
Recall status
Terminated
CDRH record ID
187825
FDA record created
2021-06-17
Recall event number
88075

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗