Apheresis Technologies, Inc. is taking the following Corrective Action: Modification of the Owners Manual Only for the PlasmaPro Plasma Pump Model PP-05. Operator's Manual will modify tubeset loading instructions so as to reduce possible loading error which might result in the tube being pulled into the pump head.
Source facts
Action
PlasmaPro Plasma Pump Operators Manual was modified and distributed to firms consignees. Firms received written notification of re-issuance of Operator's Manual.
Initiated
2009-04-01
Root cause
Labeling design
Terminated
2010-07-06
Distribution
Nationwide Distribution -- Florida, Colorado and Utah.
Recall number
Z-1897-2009
Recall status
Terminated
CDRH record ID
83535
FDA classified
2009-08-27
Recall event number
52640