Open original FDA record ↗Recall record
- FDA identifier
- Z-1897-2010
- Company in FDA record
- Haag-Streit USA Inc
- FDA classification date
- Jun 24, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During the creation of the Lenstar LS 900 software, the ability to perform a manual adjustment of the retinal thickness measurement was not disabled for products targeted for US distribution. The current software installed allows the user to manually alter the retinal thickness measurement, which could modify the automatic axial eye length measurement resulting in the selection of an incorrect in
Source facts
- Action
- HAAG - Streit USA, Inc. issued notification letters to their customers informing them of the problem. They instructed their customer to refrain from adjusting the gates for the measurement of retinal thickness until after the system has been updated. For further information, contact HAAG - Streit USA, Inc. at 1-513-336-7255.
- Initiated
- 2010-05-19
- Root cause
- PMA
- Terminated
- 2012-02-24
- Distribution
- Nationwide Distribution -- Including AL, AZ, CA, FL, ID, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, VA, VT, WA and WV.
- Recall number
- Z-1897-2010
- Recall status
- Terminated
- CDRH record ID
- 92075
- FDA classified
- 2010-06-24
- Recall event number
- 55857