- FDA identifier
- Z-1897-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jun 28, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Source facts
- Action
- Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.
- Initiated
- 2019-06-06
- Root cause
- Device Design
- Terminated
- 2021-02-09
- Distribution
- Nationwide
- Recall number
- Z-1897-2019
- Recall status
- Terminated
- CDRH record ID
- 173463
- FDA record created
- 2019-06-28
- Recall event number
- 83082