Z-1897-2020: da Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of t

Recall record Z-1897-2020 for Intuitive Surgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1897-2020
Company in FDA record
Intuitive Surgical Inc
FDA record creation date
May 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.

Source facts

Action
On 04/11/2019, the firm sent an "Urgent Medical Device Correction" notifications via FedEx to its customers informing them that the Recalling Firm has become aware that specific Surgical Systems arms have the potential to encounter friction in specific instrument arm orientations. As a result, sluggish, jerky, or hesitating motion may be experienced at the instrument tip and/or resistance felt in the hand controls. To correct this issue, a Recalling Firm's representative will schedule a site visit to perform evaluations on their systems. Based on the results of the evaluation, any necessary corrections will be prioritized appropriately. In addition: Customers are informed and instructed that If any of instruments arms exhibit sluggish, jerky, or hesitating motion, or if you feel significant resistance from the hand controls with no interference between arms/instruments, contact the da Vinci Surgical Technical Assistance Team (dVSTAT) as soon as possible. General Acknowledgement 1. Forward this letter to your Risk Manager, OR Director, Biomedical Engineering staff, and other members of your medical staff who perform da Vinci procedures. 2. If available, please log into the da Vinci Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 3. In the event the da Vinci online resource cannot be used, complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form. 4. Please retain a copy of this letter and the acknowledgement form for your files. representatives will be available by phone to answer any questions related to this Medical Device Correction. An Intuitive representative will schedule a site visit with your hospital to perform the inspection and provide any potential correction activity for the instrument arms of your system(s). If you need further information or support concerning this Medical Devic
Initiated
2019-04-11
Root cause
Process control
Terminated
2024-04-08
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.
Recall number
Z-1897-2020
Recall status
Terminated
CDRH record ID
179353
FDA record created
2020-05-05
Recall event number
83482

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗