Skip to content

Public FDA record

Z-1898-2021: VENOVO Venous Stent System 8F Product Description(REF)/dimension: VENEL10040 /Venovo 8F 10/40/1200mm OUS

Recall record Z-1898-2021 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1898-2021
Company in FDA record
Bard Peripheral Vascular Inc
FDA record creation date
Jun 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.

Source facts

Action
On 01/13/2021, the firm sent an "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the Recalling Firm has received reports regarding the venous stent system 9F where the proximal end of the stent does not immediately expand upon deployment, and remains connected to the delivery system, and that over-manipulation or forcing the catheter delivery system in attempts to assist the stent's expansion, could potentially have a varying degree of harm associated with it. potential harm ranges from prolonging the procedure, damaging or deformity of the stent, potential vascular injury and / or hemodynamic disruption affecting the blood flow and / or a thrombotic event. The Recalling Firm is informing customers that the affected product can continue to be used, but are instructing them to: 1. Share this communication with all physicians using the Venovo" Venous Stent Systems product within their organization to ensure that they are aware of this Safety Notice. 2. Continue the use of these Venovo" Venous Stent Systems 9F, 14mm products following the Safety Notice guidance. 3. Report any adverse health consequences experienced with the use of this product to the Recalling Firm. A list of the affected Product Codes and Lot/Serial Numbers are included as Appendix 1 attached to the Recall Notification Letter. For questions/inquiries/assistance contact the Recalling Firm's Customer/Technical Support at 1-888-237-2762 (Opt 4, Opt 2) Monday thru Friday 8:00 a.m. - 5:00 p.m. CST Updated Information for Recall Expansion (dated 05/11/2021): On 05/07/2021, the firm sent additional " "URGENT MEDICAL DEVICE SAFETY NOTICE" via FedEx to customers informing them that the scope of the recall (initiated on 01/13/2021) was being expanded to include all sizes and lots within expiry date. Customer are instructed to: 1. Check all inventory locations within their institution for the affected Venous Stent System listed in Attachment A and immediately
Initiated
2020-12-14
Root cause
Under Investigation by firm
Distribution
Worldwide distribution. US Nationwide, ALGERIA, ANDORRA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BRUNEI, BULGARIA, CHILE, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, DEUTSCHLAND, EGYPT, FINLAND, FRANCE, GEORGIA, GREECE, HONG KONG, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, MALAYSIA, MEXICO, NETHERLANDS, NORWAY, OMAN, PANAMA, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDIA ARABIA, SINGAPORE, SLOVAKIA (SLOVAK REPUBLIC), SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TURKEY, UNITED ARAB EMIRATES (UAE), UNITED KINGDOM, VIETNAM
Recall number
Z-1898-2021
Recall status
Open, Classified
CDRH record ID
187685
FDA record created
2021-06-17
Recall event number
87208

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗