Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
Source facts
Action
FUJIFILM Healthcare Americas Corporation (FUJIFILM) issued Urgent -FIrst Notice letter via email on 4/11/24. Letter states reason for recall, health risk and action to take:
1)Please complete and return the form attached on page 3 of this communication. Providingthe response with the information requested is essential for ensuring appropriate action istaken
2)Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a serviceticket.
3)Until your upgrade is scheduled and completed, you can continue to safely use FDRVisionary Suite, however, inspect the bolts prior to use.
Please contact our field action coordinator, Chaitrali Kulkarni, via email at hcusregulatoryaffairs@fujifilm.com, if you have any further questions regarding this field action.
Initiated
2024-04-11
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Recall number
Z-1898-2024
Recall status
Open, Classified
CDRH record ID
207224
FDA classified
2024-05-24
Recall event number
94443