Z-1898-2024: FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spin

Recall record Z-1898-2024 for FUJIFILM Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1898-2024
Company in FDA record
FUJIFILM Healthcare Americas Corporation
FDA classification date
May 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Source facts

Action
FUJIFILM Healthcare Americas Corporation (FUJIFILM) issued Urgent -FIrst Notice letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: 1)Please complete and return the form attached on page 3 of this communication. Providingthe response with the information requested is essential for ensuring appropriate action istaken 2)Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a serviceticket. 3)Until your upgrade is scheduled and completed, you can continue to safely use FDRVisionary Suite, however, inspect the bolts prior to use. Please contact our field action coordinator, Chaitrali Kulkarni, via email at hcusregulatoryaffairs@fujifilm.com, if you have any further questions regarding this field action.
Initiated
2024-04-11
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.
Recall number
Z-1898-2024
Recall status
Open, Classified
CDRH record ID
207224
FDA classified
2024-05-24
Recall event number
94443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗