Z-1899-2008: Model 6000 CMS, CMS IOD, or Plus Ambulatory Infusion Pump, Model Number: 360-1101

Recall record Z-1899-2008 for Curlin Medical Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1899-2008
Company in FDA record
Curlin Medical Llc
FDA classification date
Sep 22, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Curlin Medical has investigated five cases of over-delivery/free-flow in the past eight months involving Curlin ambulatory peristaltic multi-therapy infusion pumps where the pumping chamber door (platen) had become deformed in a manner that permitted gravity flow. The firm's investigations have identified the cause of this deformity to have been a result of the pump being dropped with the door (p

Source facts

Action
On May 30, 2008, Curlin Medical began distribution of a "Voluntary Field Correction Advisory Notice" to all customers in the United States via USPS Registered Mail. The Notice informed customers of the problem and gave the customers (1) Guidelines on testing pump volumetric accuracy, (2) Updated user manuals to elevate the "caution" statement related to dropped pumps to a "warning" statement, and (3) A warning label to be located on the pump: "Warning: Impact may cause damage. If dropped, pump must be checked for accuracy prior to use". A toll free number (888-287-5999) was given for customers experiencing issues that require immediate assistance.
Initiated
2008-05-30
Root cause
Other
Terminated
2009-06-11
Distribution
Worldwide Distribution --- including USA and countries of Australia, Croatia, Germany, Italy, New Zealand, and South Africa.
Recall number
Z-1899-2008
Recall status
Terminated
CDRH record ID
71776
FDA classified
2008-09-22
Recall event number
48639

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗