Skip to content

Public FDA record

Z-1899-2009: Vertier Surgical Table, an AC powered, general purpose, mobile surgical table with electro-hydraulically

Recall record Z-1899-2009 for Stryker Communications Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1899-2009
Company in FDA record
Stryker Communications Corp
FDA classification date
Aug 26, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Surgical tables were produced using an unapproved tilt frame bushing.

Source facts

Action
Stryker Communications issued an "Urgent: Medical Device Recall" notice dated November 13, 2008 informing Consignees of the affected product. The firm intends to contact each facility to coordinate service by a trained representative. For further information, contact Stryker at 1-972-410-7100.
Initiated
2008-11-13
Root cause
Nonconforming Material/Component
Terminated
2009-08-28
Distribution
Worldwide Distribution -- United States (OR, IL, WA, CA, AR, TX, UT, SC and GA) and Canada.
Recall number
Z-1899-2009
Recall status
Terminated
CDRH record ID
83919
FDA classified
2009-08-26
Recall event number
52712

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗