Open original FDA record ↗Recall record
- FDA identifier
- Z-1899-2017
- Company in FDA record
- Roche Diabetes Care, Inc.
- FDA record creation date
- May 1, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
Source facts
- Action
- Roche Diabetes Care, Inc. discovered a software issue in the Accu-Chek Connect app versions 2.0.0, 2.0.1, and 2.1.0 that can result in incorrect bolus recommendations. This software bug may result in errors in the active insulin calculation, creating the potential for a false high insulin bolus recommendation. An insulin bolus delivered based on potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Customer notices were distributed via mail on 03/30/2017, and the firm's website was also updated with recall information on that day. Customers were instructed to do the following: discontinue use of the bolus advisor feature in app versions 2.0.0, 2.0.1, and 2.1.0 (iOS & Android) until the corrected version is available. Provide patients with a copy of the enclosed Urgent Medical Device Correction 17-001. Complete the enclosed faxback form and fax it to Roche Diabetes Care, Inc., at 1-888-912-8456. If you have questions about the information in this notification, please contact Accu-Chek Customer Care by visiting accu-chek.com or by calling 1-800-688-4578. The corrected version of the app is expected to be available on or about April 13, 2017. New patients will be able to begin using the bolus advisor feature once the corrected version is available. Please advise your patients to visit accu-chek.com for updates regarding the availability of the corrected version of the app.
- Initiated
- 2017-03-14
- Root cause
- Software design
- Terminated
- 2018-01-16
- Distribution
- US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, DE, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
- Recall number
- Z-1899-2017
- Recall status
- Terminated
- CDRH record ID
- 154435
- FDA record created
- 2017-05-01
- Recall event number
- 76870